GUI Reference
R&S
®
ZNB/ZNBT
492
User Manual 1173.9163.02 ─ 38
Example:
A mathematical trace value amounts to 1 (real value); the port impedance is 50 Ω. If
"Result is Wave Quantity" is on, the analyzer assumes the trace value to be 1 V, which
is converted into a linear power of 20 mW, corresponding to approx. 13 dBm. With
"Result is Wave Quantity" off, the trace value 1 is directly converted into a logarithmic
power of 0 dBm.
Tip:
See also example for
CALCulate<Chn>:MATH:WUNit:STATe
.
Remote command:
CALCulate<Chn>:MATH:WUNit[:STATe]
Recall... / Save...
Recalls / saves a mathematical expression from / to a trace math string file. Trace
math string files are ASCII files with the default extension
*.mth
and contain the math-
ematical expression as it is written in the "User Def Math" dialog. It is possible to
change or create math string files using a text editor.
Remote command:
CALCulate<Chn>:MATH:WUNit[:STATe]
5.12.3 All Mem All Data Tab
Performs actions on all data or memory traces in the active recall set.
Background information
Refer to
All Data to <Destination>
Stores the current data of all data traces in the active recall set to memory traces, in
accordance with the
setting. No trace functions are applied to the stored
traces.
Trace Config Softtool