GUI Reference
R&S
®
ZNB/ZNBT
462
User Manual 1173.9163.02 ─ 38
Imb21 / Imb12 / CMRR21 / CMRR12
Selects one of the standard 2-port
parameters as a measured
quantity for the active trace. The buttons are available if either logical port 1 or logical
port 2 (or both) is defined as a balanced port. A yellow font color indicates that the
other logical port number is currently undefined and the physical port with the same
number is marked as unused; it will be enabled (as single-ended port with "logical port
number" = "physical port number") for the measurement.
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "IMB21" |
"IMB12" | "CMRR21" | "CMRR12"
Create new trace and select trace name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "IMB21" |
"IMB12" | "CMRR21" | "CMRR12"
Balanced Ports...
Opens a dialog to define a balanced port configuration.
See
Chapter 5.9.1.3, "Balanced Ports Dialog"
5.9.8.2
Imbalance Differential Dialog
Allows to measure the differential mode imbalance of a DUT with balanced input and
output ports, connected to a balanced test port and two single-ended test ports
Access:
TRACE – MEAS > "Imbal. CMRR" > "Imbalance Differential..."
●
The
Logical Port
on the left represents the balanced test port: any (active) bal-
anced logical port can be selected
●
Physical Port 1 and Physical Port 2 represent the two single-ended test ports: any
two (active) single-ended logical ports can be selected
●
The part below allows to select the signal direction and hence the "Imbalance
Result" to be calculated.
Meas Softtool