GUI Reference
R&S
®
ZNB/ZNBT
437
User Manual 1173.9163.02 ─ 38
S-Parameter
as a measured quantity for the active trace.
Single-ended (unbalanced) S-parameters are referred to as S<out>< in>, where <out>
and <in> denote the output and input
logical
port numbers, respectively.
If entered manually, <out> and <in> must be expressed using the same number of dig-
its, i.e. for <out>=10 and <in>=9 you may specify S<out>< in> as S1009 or S010009,
but not as S109. For a channel setup with n single-ended ports, n
2
single-ended S-
parameters can be measured.
In presence of balanced ports, standard S-parameters are defined in the form
S<m_out><m_in><out><in>, where output mode <m_out> and input mode <m_in> can
be one of
●
d (differential, balanced)
●
c (common, balanced)
●
s (single-ended, unbalanced)
It is also possible to display "raw" single-ended S-parameters within the same channel:
after setting the S-Parameter type to "S" any pair of (used)
physical
ports may be
selected.
Remote command:
CALCulate<Ch>:PARameter:MEASure
CALCulate<Ch>:PARameter:MEASure:SENDed
CALCulate<Ch>:PARameter:SDEFine
CALCulate<Ch>:PARameter:SDEFine:SENDed
S<out><in> (softkeys)
Selects one of the four elements of the standard 2-port
quantity for the active trace.
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "S11" | "S12" |
"S21" | "S22"
Create new trace and select trace name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "S11" | "S12" |
"S21" | "S22"
All S-Params
Creates n
2
diagrams (for an n-port vector network analyzer) and displays the full set of
, one in each diagram. The diagrams are arranged as an (n x n) matrix.
The reflection coefficients S
ii
appear in the diagrams on the main diagonal, the trans-
mission coefficients S
ij
(i
≠
j) occupy the other diagrams. Reflection coefficients are dis-
played in Smith diagrams; transmission coefficients in Cartesian diagrams with loga-
rithmic ("dB mag") scale.
See also
.
Note:
This function is disabled if the number of used test ports is greater than 9.
Remote command:
CALCulate<Ch>:PARameter:SDEFine
CALCulate<Ch>:PARameter:DEFine:SGRoup