GUI Reference
R&S
®
ZNB/ZNBT
368
User Manual 1173.9163.02 ─ 38
Figure 5-28: Define Parallel Measurement (R&S
ZNBT8 with 24 ports)
●
Modified settings are not applied unless the "Define Parallel Measurement" dialog
is closed with the "OK" button.
●
On "OK" any preexisting logical port configuration is overwritten.
DUT Definition Tab
Auto Setting (R&S
ZNBT only)
The "Auto Setting" functionality allows to quickly set up a parallel measurement of mul-
tiple identical DUTs, assuming the DUTs are fully connected in an aligned way, i.e. port
j
∈
{1,...,"Port Count"}
of DUT
i
∈
{1,...,"Number of DUTs"}
is connected to VNA port
(i-1) · "Port Count" + j
.
In particular this requires
"Number of DUTs" · "Port Count"
to be less or equal than the
number of available physical VNA ports.
For instance, on a 16-port R&S
ZNBT you can set up 4 four-port DUTs in a few sec-
onds:
1. Set "Number of DUTs" and "Port Count" to 4
2. Use "Set" to prepare the DUT configuration
Modify the DUT/connection configuration according to your needs (see procedures
below).
Channel Config Softtool