GUI Reference
R&S
®
ZNB/ZNBT
315
User Manual 1173.9163.02 ─ 38
Background and related information
●
The "Multiple Cal in Calibration Wizard" feature is disabled by default and has to be
activated in the
tab of the "System Config" dialog.
●
Chapter 5.4.1.9, "SMARTerCal Wizard (Manual)"
ter 5.4.1.10, "SMARTerCal Wizard (Manual, MultiCal) "
on page 314 for more
information.
Step 1: Ports and Type
The first page of the wizard presents the list of already configured SMARTERCals and
allows to
●
"Add" new SMARTerCals
●
"Delete" or "Modify" existing SMARTerCals
Access:
CHANNEL > CAL > Start Cal > SMARTerCal
–
Start... (Manual)
Calibrate all Channels
Check this box to apply the acquired correction data to all channels in the active recall
set. Leave it unchecked (preset setting) to apply them only to the active channel.
Note that this option is available only if the active recall set contains multiple channels.
Remote command:
[SENSe:]CORRection:COLLect:CHANnels:ALL
Cal Softtool