GUI Reference
R&S
®
ZNB/ZNBT
270
User Manual 1173.9163.02 ─ 38
2.
"Connections":
Define the port assignment(s) between the R&S
ZNB/ZNBT and
the calibration unit(s).
3.
"Cal Unit":
Acquire measurement data for the individual port assignment(s) and
the necessary measurement standards (provided by the calibration unit). Calculate
the system error correction data (error terms) from the measurement data and
apply the result to the active channel.
Background and related information
●
The "Multiple Cal in Calibration Wizard" feature is disabled by default and has to be
activated in the
of the System Config dialog.
●
Chapter 5.4.1.2, "Calibration Unit Wizard"
on page 260 for further infor-
mation.
Step 1: Ports
The first page of the wizard presents the list of already configured calibrations and
allows to
●
"Add" new calibrations
●
"Delete" or "Modify" existing calibrations
Calibrate all Channels
Check this box to apply the acquired correction data to all channels in the active recall
set. Leave it unchecked (preset setting) to apply them only to the active channel.
Note that this option is available only if the active recall set contains multiple channels.
Remote command:
[SENSe:]CORRection:COLLect:CHANnels:ALL
Add...
Adds a new system error correction.
The "Add" button opens the
without pre-selected ports.
Cal Softtool