Concepts and Features
R&S
®
ZNB/ZNBT
167
User Manual 1173.9163.02 ─ 38
Calibration of S-parameters
S-parameters are not affected by a scalar power calibration. S-parameters are ratios of
incident and outgoing waves: for linear DUTs, they do not depend on the absolute
power. For measurements on non-linear DUTs, a SMARTerCal is recommended.
A SMARTerCal is also appropriate for frequency conversion measurements. For
detailed information refer to
Table 4-11: Recommended calibration methods for various measurements
Measurement
System error correction Scalar Power calibration
SMARTerCal
S-parameter meas. on linear
DUTs
Yes
Not necessary
Not necessary
S-parameter meas. on non-lin-
ear DUTs
Not possible
--> Use SMARTerCal
Yes
Power (source) and meas. receiver, sup-
ported in combination with SMARTerCal
Yes
1)
Meas. of wave quantities or
ratios on linear DUTs
Not possible
--> Use SMARTerCal
Not necessary
Yes
Meas. of wave quantities or
ratios on non-linear DUTs
Not possible
--> Use SMARTerCal
Yes
Power (source) and meas. receiver, sup-
ported in combination with SMARTerCal
Yes
Power sweep, e.g. for com-
pression point measurement
Yes
Power (source): Yes
Meas. receiver: not necessary
Not necessary
Frequency conversion mea-
surements on linear DUTs
Not possible
--> Use SMARTerCal
Not necessary
Yes
Frequency conversion mea-
surements on non-linear DUTs
Not possible
--> Use SMARTerCal
Yes
Power (source) and meas. receiver, sup-
ported in combination with SMARTerCal
Yes
1) SMARTerCal is a multi-port calibration type. For one-port measurements, perform a
two-port SMARTerCal for a port pair including the measurement port.
In general, a power calibration involves two stages:
1.
Source power calibration:
An external power meter is connected to the calibra-
tion plane. The analyzer uses the power meter readings to calibrate its reference
receiver. Subsequently, it modifies its source power so that the calibrated refer-
ence receiver reading corresponds to the desired source power value (flatness cal-
ibration).
2.
Measurement receiver calibration:
The analyzer uses the calibrated source sig-
nal to adjust the power reading at the receive port.
Calibration