Command Reference
R&S
®
ZNB/ZNBT
1084
User Manual 1173.9163.02 ─ 38
Example:
*RST
Reset the instrument, activating a frequency sweep with 201
sweep points.
SWE:TIME?
Query total sweep time.
SWE:POIN 2010
Multiply the (default) number of points by 10.
SWE:TIME?
Query total sweep time again. The analyzer estimates a sweep
time that is also multiplied by 10.
Manual operation:
See
[SENSe<Ch>:]SWEep:STEP
<StepSize>
Sets the distance between two consecutive sweep points in a (non-segmented) linear
frequency sweep.
Suffix:
<Ch>
.
Channel number
Parameters:
<StepSize>
Frequency step size.
The step size is equal to the current sweep spandivided by the
number of sweep points minus one (see
and
,
respectively). See also the description of manual control and the
program example below.
Range:
Depends on the current sweep span and the maxi-
mum number of sweep points: "Span Frequency" /
(max. "Number of Points" – 1) ≤ "Freq Step Size" ≤
"Span Frequency"
*RST:
Depends on the analyzer model. The default step
size is equal to the default sweep span of the ana-
lyzer divided by the default number of sweep points
minus one.
Default unit: Hz
Example:
*RST; :SWE:STEP?
Query the default step size. Currently for all analyzers the
default sweep span is the full frequency range and the default
number of points is 201. Hence the response is (max. frequency
- min. frequency) / 200.
SWE:STEP UP
Increase the step size.
FREQ:STOP?; :SWE:POIN?
Query the stop frequency of the sweep and the number of
points. Increasing the step size has changed both values.
Manual operation:
See
SCPI Command Reference