Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
86
User Manual 1179.3315.02 ─ 01
Suffix:
<ip>
.
1 | 2
For R&S
FSW85 models with two RF input connectors:
1: Input 1 (1
mm [RF Input] connector)
2: Input 2 (1.85
mm [RF2 Input] connector)
For all other models:
irrelevant
Setting parameters:
<DeviceName>
string
Name of the instrument for which the RF input protection is to be
reset.
Example:
INP:ATT:PROT:RES 'MyDevice'
Manual operation:
See
INPut<ip>:COUPling<ant>
<CouplingType>
This command selects the coupling type of the RF input.
Suffix:
<ip>
.
1 | 2
irrelevant
<ant>
(for MIMO measurements only)
Parameters:
<CouplingType>
AC | DC
AC
AC coupling
DC
DC coupling
*RST:
AC
Example:
INP:COUP DC
Manual operation:
See
INPut<ip>:DPATh
<DirectPath>
Enables or disables the use of the direct path for frequencies close to 0
Hz.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<DirectPath>
AUTO | OFF
AUTO | 1
(Default) the direct path is used automatically for frequencies
close to 0
Hz.
Configuring the Measurement