Preparing for Use
R&S
®
TSME
28
User Manual 1514.6542.02 ─ 10
Figure 2-6: R&S
TSME LTE MIMO 4x4 setup with synchronization cable
Further connections for setups with multiple R&S
TSMEs
●
Connect all R&S
TSMEs to a power supply. Optionally, all devices can be connec-
ted to the same power supply.
The R&S
TSME-ZYC option provides a cable to connect 2 R&S
TSMEs to one
power source
The R&S
TSME-ZYC4 option provides a cable to connect four R&S
TSMEs to one
power source, such as the R&S TSMW-Z1 AC power supply.
●
Connect an RF antenna (not included in package) to the R&S
TSME.
●
You only have to connect a GPS antenna to one of the connected R&S
TSMEs. It
is recommended that you connect it to the R&S
TSME with the lowest IP address.
●
Configure the LAN connection to the host PC as described in the following chapter.
2.2.5.5
Configuring the LAN Connections Between Multiple R&S
TSMEs Connected to
One Host PC
The LAN connection from multiple R&S
TSMEs to the same host PC can be implemen-
ted using two different methods:
●
Using separate network interface connectors (NICs)
Connecting the R&S
TSME