for lack of space, slots cannot remain free between modules with DC/DC converter
module R&S TS-PDC in the rear IO area, or
●
the R&S TS-PWA3 is not operated in normal horizontal standard position.
Heat generation
Adequate space for air inlet and outlet should be allowed when the case is rack moun-
ted, with at least half a height unit (22 mm) being left above and below the chassis!
For desktop mounting, ensure that the case is not covered over!
3.3.11
Option R&S TS-PK04P
If the optional R&S TS-PK04P (1157.9104.03) cable set has been installed in the chas-
sis, some channels of suitable switch modules (R&S TS-PSM1, R&S TS-PSM3, R&S
TS-PSM4 und R&S TS-PSM5) will be made accessible via the connectors X3 and X4
on the rear side of the chassis. For this purpose, the switch module must be installed in
slot 16.
Figure 3-15: R&S TS-PSM1 with equivalent cable set in a CompactTSVP
With installed option R&S TS-PK04P, the switch modules R&S TS-PMB and TS-PSM2
cannot be operated in slot 16.
3.4
System Module R&S TS-PSYS2
3.4.1
General
The R&S TS-PSYS2 is in the
Rear I/O slot 15
of the R&S TS-PWA3.
System functions such as voltage and temperature monitoring, switchable trigger sig-
nals and optocoupler interface are used to integrate the PowerTSVP in a complete
system.
The module is available in two variants:
●
TS-PSYS2 1157.8508.02 (obsolete)
System Module R&S TS-PSYS2