![Rohde & Schwarz SMM100A User Manual Download Page 962](http://html.mh-extra.com/html/rohde-and-schwarz/smm100a/smm100a_user-manual_1477939962.webp)
Remote control commands
R&S
®
SMM100A
962
User Manual 1179.1341.02 ─ 05
Example: Restart network
SYSTem:COMMunicate:NETWork:RESTart
// terminates the network connection and sets it up again
Example: Finding out the used VISA resource strings
SYSTem:COMMunicate:NETWork:RESource?
// "TCPIP::10.113.0.104::inst0::INSTR"
SYSTem:COMMunicate:HISLip:RESource?
// "TCPIP::10.113.0.104::hislip0::INSTR"
SYSTem:COMMunicate:SOCKet:RESource?
// "TCPIP::10.113.0.104::5025::SOCKET"
SYSTem:COMMunicate:USB:RESource?
// "USB::0x0AAD::0x0277::100001::INSTR"
SYSTem:COMMunicate:GPIB:RESource?
// "GPIB::28::INSTR"
SYSTem:COMMunicate:GPIB:SELF:ADDRess?
// 28
SYSTem:COMMunicate:GPIB:LTERminator?
// STAN
SYSTem:COMMunicate:SERial:RESource?
// "ASRL1::INSTR"
SYSTem:COMMunicate:SERial:SBITs?
// 1
SYSTem:COMMunicate:SERial:BAUD?
// 115200
SYSTem:COMMunicate:SERial:PARity?
// NONE
Example: Querying the error queue
SYSTem:ERRor:STATic?
// -221,"Settings conflict", 153,"Input voltage out of range", ...
// returns all static errors that are collected in the error queue
SYSTem:ERRor:HISTory:ClEar
// deletes the history entries
......................................................................................... 964
.....................................................................................965
............................................................................................... 965
............................................................................................... 966
............................................................................................... 966
......................................................................................966
............................................................................................... 967
SYSTem subsystem