Remote control commands
R&S
®
SMM100A
958
User Manual 1179.1341.02 ─ 05
Parameters:
<CombFreqStop>
float
Range:
-59999E5 to 12E9
Increment: 0.01
*RST:
500E6
Example:
See
Example "Setup an RF frequency or power sweep"
[:SOURce<hw>]:COMBined:POWer:STARt
<CombPowStart>
Sets the start level value of the combined RF frequency / level sweep.
See
Chapter 7.10.2.1, "Correlating parameters in sweep mode"
Parameters:
<CombPowStart>
float
Range:
-245 to 120
Increment: 0.01
*RST:
-30
Example:
See
Example "Setup an RF frequency or power sweep"
[:SOURce<hw>]:COMBined:POWer:STOP
<CombPowStop>
Sets the stop level value of the combined RF frequency / level sweep.
Parameters:
<CombPowStop>
float
Range:
-245 to 120
Increment: 0.01
*RST:
-10
Example:
See
Example "Setup an RF frequency or power sweep"
[:SOURce<hw>]:SWEep:COMBined:COUNt
<StepCount>
Defines the number of sweeps you want to execute.
This parameter applies to
[:SOURce<hw>]:SWEep:COMBined:MODE > SINGle
. To
start the sweep signal generation, use the command
Parameters:
<StepCount>
integer
Range:
1 to SeMAX_INT_STEP-1
*RST:
501
Example:
See
Example "Setup an RF frequency or power sweep"
SOURce subsystem