Remote control commands
R&S
®
SMM100A
907
User Manual 1179.1341.02 ─ 05
[:SOURce]:LFOutput<ch>[:STATe]
<State>
Activates LF signal output.
Parameters:
<State>
1 | ON | 0 | OFF
*RST:
0
Example:
See
Example "Configuring the LF generator"
Manual operation:
See
[:SOURce]:LFOutput<ch>:OFFSet
<Offset>
Sets a DC offset at the selected LF Output.
Parameters:
<Offset>
float
Range:
-3.6 to 3.6
Increment: 2E-3
*RST:
0
Default unit: V
Example:
See
Example "Configuring the LF generator"
Manual operation:
See
[:SOURce]:LFOutput<ch>:SOURce
<Source>
Determines the LF signal to be synchronized, when monitoring is enabled.
Parameters:
<Source>
LF1 | LF2 | NOISe | AM | FMPM | EXT1
*RST:
LF1
Example:
See
Example "Configuring the LF generator"
Manual operation:
See
[:SOURce]:LFOutput<ch>:SOURce:PATH
<SourPath>
Determines the path of the LF output source.
Parameters:
<SourPath>
A | B
*RST:
A
Example:
LFO:SOUR:PATH?
Queries the currently set path for the LF output signal source.
[:SOURce]:LFOutput<ch>:VOLTage
<Voltage>
Sets the output voltage of the selected LF output.
SOURce subsystem