Remote control commands
R&S
®
SMM100A
812
User Manual 1179.1341.02 ─ 05
<Digits>
Decimal value
Gives the number of decimal digits used for the
<Length>
value
<Length>
Decimal value
Number of bytes the follow in the
<Binary data>
part
<Binary data>
Binary data in ASCII format
Query parameters:
<Tag>
'comment' | 'copyright' | 'date' | 'lacpfilter' | 'marker name' |
'poweroffset'
Example:
Query
See
Example "Managing waveform files"
Example:
Setting
SOURce:BB:ARB:WAVeform:DATA "/var/user/test1.wv",#220<binary data>
// Writes the binary block data <binary data> to file test1.wv
// <binary data> contains 20 bytes
// <binary data> is a placeholder;
// the actual ASCII values are not printable
[:SOURce<hw>]:BB:ARBitrary:WAVeform:TAG?
Queries the content of the specified tag of the selected waveform file (see also
ter 4.6.5, "Tags for waveforms, data and control lists"
Return values:
<Tag>
'comment' | 'copyright' | 'date' | 'lacpfilter' | 'marker name' |
'poweroffset' | 'samples'
Example:
Example "Managing waveform files"
Usage:
Query only
[:SOURce<hw>]:BB:ARBitrary:CLOCk
<Clock>
Sets the clock frequency.
If you load a waveform, the clock rate is determined as defined with the waveform tag
. This command subsequently changes the clock rate; see
data sheet for value range.
Parameters:
<Clock>
float
Range:
depends on the installed options
Increment: 0.001
*RST:
1E6
Default unit: Hz
Example:
See
Example "Managing waveform files"
SOURce subsystem