Remote control commands
R&S
®
SMM100A
704
User Manual 1179.1341.02 ─ 05
:SLISt:CLEar[:ALL]
Removes all R&S NRP power sensors from the list.
Example:
See
Example "Detecting and assigning a power sensor"
Usage:
Event
Manual operation:
See
:SLISt:ELEMent<ch>:MAPPing
<Mapping>
to one of the four sensor channels.
Parameters:
<Mapping>
SENS1 | SENSor1 | SENS2 | SENSor2 | SENS3 | SENSor3 |
SENS4 | SENSor4 | UNMapped
Sensor channel.
*RST:
UNMapped
Example:
See
Example "Detecting and assigning a power sensor"
Manual operation:
See
:SLISt:SENSor:MAP
<SensorId>, <Mapping>
Assigns a sensor directly to one of the sensor channels, using the sensor name and
serial number.
To find out the the sensor name and ID, you can get it from the label of the R&S NRP,
or using the command
. This command detects all R&S NRP
power sensors connected in the LAN or via 'USBTMC protocol.
Setting parameters:
<SensorId>
string
<Mapping>
enum
Example:
See
Example "Detecting and assigning a power sensor"
Usage:
Setting only
Manual operation:
See
:INITiate<hw>[:POWer]:CONTinuous
<Continuous>
Switches the local state of the continuous power measurement by R&S NRP power
sensors on and off. Switching off local state enhances the measurement performance
during remote control.
SENSe, READ, INITiate and SLISt subsystems