Troubleshooting and notifications
R&S
®
SMM100A
1000
User Manual 1179.1341.02 ─ 05
The message is displayed until the error is eliminated. It covers the status display in
the info line. After error elimination, the message automatically disappears and is also
recorded in the history.
Remote command:
14.2
SCPI notifications
The SCPI notifications are similar for all SCPI instruments. Detailed information and an
overview of all notifications as defined in SCPI standard can be found in the corre-
sponding documentation.
SCPI notifications have negative codes (numbers). The error text being entered into
the error/event queue or being displayed is printed in bold face on the left together with
the error code. Below the error text, there is an explanation to the respective error.
14.3
Device-specific notifications
The following table contains all error messages specific for the instrument in alphabeti-
cal order and an explanation of the error situation. The positive error codes mark the
errors specific to the instrument.
The device-specific error messages set bit 3 in the ESR register.
The index provides a list of the error messages sorted according to their error codes.
Error
code
Error
Description
Remedy
50
Extern reference out of
range or disconnected
External reference is selected but no external sig-
nal is applied or the signal is out of range.
●
Check the selected reference signal
source (internal or external) in the
"Setup > Reference Oscillator" dia-
log.
●
Change setting to 'internal' if no
appropriate external source is avail-
able.
140
This modulation forces
other modulations off
A modulation has been switched on which cannot
be used at the same time as an already active
modulation. The previous modulation has been
switched off.
Example: Enabling modulation GSM/EDGE
switches off any active digital modulation.
180
Adjustment failed
Adjustment could not be executed.
Generate the adjustment data and load
it into the device
182
Adjustment data missing
Adjustment data is missing.
Generate the adjustment data and load
it into the device
Device-specific notifications