![Rohde & Schwarz SMF-B1 Operating Manual Download Page 573](http://html1.mh-extra.com/html/rohde-and-schwarz/smf-b1/smf-b1_operating-manual_1477740573.webp)
Remote Control Commands
R&S
®
SMF100A
558
Operating Manual 1167.2319.02 ─ 12
[:SOURce<hw>]:SWEep[:FREQuency]:MARKer:AOFF
The command switches off all frequency markers. The command triggers an event and
therefore has no query form and no *RST value.
Example:
SOUR:SWE:FREQ:MARK:AOFF
switches off all frequency markers.
Usage:
Event
[:SOURce<hw>]:SWEep[:FREQuency]:MARKer<ch>:FREQuency
<Frequency>
Sets the frequency of the selected marker. If the marker is switched on, the signal level
at the MARKER output changes on reaching the frequency mark.
Parameters:
<Frequency>
float
Increment: 0.01
*RST:
100 + n* 100
Example:
SWE:FREQ:MARK1:FREQ 1.5GHz
sets the frequency mark for marker 1 to 1.5 GHz.
Manual operation:
See
" Frequency - RF Frequency Marker "
[:SOURce<hw>]:SWEep[:FREQuency]:MARKer<ch>:FSTate
<FState>
The command switches on or off the selected marker. If the marker is switched on, the
signal level at the MARKER output changes on reaching the frequency mark.
Parameters:
<FState>
0 | 1 | OFF | ON
*RST:
OFF
Example:
SWE:MARK1:FST ON
switches on marker 1.
Manual operation:
See
" State - RF Frequency Marker "
[:SOURce<hw>]:SWEep[:FREQuency]:MARKer<ch>:PSTate
<PState>
Activates or deactivates the selected amplitude marker. If activated, the level is
reduced by the amplitude entered with command
[SOURce:]SWEep[:FREQuency]:MARKer:AMPLitude
on reaching the mark.
Parameters:
<PState>
0 | 1 | OFF | ON
*RST:
OFF
Example:
SWE:MARK1:PST ON
the output level is reduced by a constant value if frequency
marker 1 is active.
SOURce Subsystem