Remote Control Commands
R&S
®
SMF100A
501
Operating Manual 1167.2319.02 ─ 12
Parameters:
<Rise>
float
Range:
1E-6 to 100
Increment: 10E-9
*RST:
0.5E-3
Example:
SOUR:LFO1:SHAP:TRI:RISE 13s
selects a rise time of 13 s for triangle shape of the signal of LF
generator 1.
Options:
R&S SMF-B20
Manual operation:
See
" Triangle Rise – LF Generator "
[:SOURce]:LFOutput<ch>:VOLTage
<Voltage>
Sers the output voltage of the LF output.
Parameters:
<Voltage>
float
Range:
0 to dynamic
Increment: 0.001
*RST:
1
Example:
SOUR:LFO1:VOLT 3V
sets the voltage of the LF output to 3 V.
Options:
R&S SMF-B20
Manual operation:
See
" Output Voltage - LF Output "
[:SOURce]:LFOutput:VOLTage:MANual
<Manual>
In Sweep mode (
:SOUR:LFO:VOLT:MODE SWE
) the command sets the level for the
next sweep step in the "Step" sweep mode (
:SOUR:LFO:SWE:VOLT:MODE MAN
).
Here only level values between the settings
[:SOUR]:LFO:VOLT:STAR
and
...:
STOP
are permitted. Each sweep step is triggered by a separate
:SOUR:LFO:VOLT:
MAN
command.
As with the "Level" value entered in the "LF Level" dialog, the "OFFSet" value is also
taken into consideration with this command.
The specified value range is therefore only effective if
:SOURce:POWer:OFFSet
is set
to 0. The value range for other "OFFSet" values can be calculated using the following
formula:
Minimum level + OFFSet ... Maximum level + OFFSet
SOURce Subsystem