Remote Control Commands
R&S
®
SMF100A
496
Operating Manual 1167.2319.02 ─ 12
Parameters:
<Shape>
SAWTooth | TRIangle
SAWTooth
One sweep runs from the start level to the stop level. The subse-
quent sweep starts at the start level again, i.e. the shape of
sweep sequence resembles a sawtooth.
TRIangle
One sweep runs from start to stop level and back, i.e. the shape
of the sweep resembles a triangle. Each subsequent sweep
starts at the start level again.
*RST:
SAWTooth
Example:
SOUR:LFO:SWE:VOLT:SHAP TRI
selects the sweep cycle with alternating ascending and
descending sweep directions.
Options:
R&S SMF-B20
Manual operation:
See
[:SOURce]:LFOutput:SWEep:VOLTage:SPACing
<Spacing>
Sets the scale function for spacing of the LF level sweep.
Parameters:
<Spacing>
LINear | LOGarithmic
*RST:
LINear
Example:
LFO:SWE:VOLT:SPAC LIN
Selects linear spacing of LV level sweep.
[:SOURce]:LFOutput:SWEep:VOLTage:STEP[:LINear]
<Linear>
Sets the step size for linear LF level sweep steps.
This parameter correlates with the number of steps
within the sweep range as follows:
f
STARt
< f
STOP
volt_points = ((f
STARt
- f
STOP
) / step_lin) + 1
Parameters:
<Linear>
float
Range:
1E-3 to 6
Increment: 1E-3
*RST:
0.1 V
Example:
SOUR:LFO:SWE:VOLT:STEP:LIN 2 or VOLT:STEP 2
sets the step width for entering the LF level to 2 V.
Options:
R&S SMF-B20
SOURce Subsystem