Remote Control Commands
R&S
®
SMF100A
493
Operating Manual 1167.2319.02 ─ 12
[:SOURce]:LFOutput:SWEep:VOLTage:MARKer:ACTive
<Active>
The command activates the selected marker. The active marker is output with an
higher voltage than all other markers.
Parameters:
<Active>
NONE | M01 | M02 | M03 | M04 | M05 | M06 | M07 | M08 | M09 |
M10
*RST:
NONE
Example:
SOUR:LFO:SWE:VOLT:MARK:ACT M05
sets marker 5 to be the active marker.
Options:
R&S SMF-B20
Manual operation:
See
" Active Marker - LF Level Marker "
[:SOURce]:LFOutput:SWEep:VOLTage:MARKer:AOFF
The command switches off all markers of level sweep. The command triggers an event
and therefore has no query form and no *RST value.
Example:
SOUR:LFO:SWE:VOLT:MARK:AOFF
switches off all level markers.
Usage:
Event
[:SOURce<hw>]:LFOutput:SWEep:VOLTage:MARKer<ch>:VOLTage
<Voltage>
The command sets the voltage mark for the selected marker. If the marker is switched
on, the signal level at the MARKER output changes on reaching the voltage mark.
Parameters:
<Voltage>
float
Range:
0 to 6
Increment: 1E-3
*RST:
0 dBm
Example:
SOUR:LFO:SWE:VOLT:MARK1:VOLT -50
sets the voltage mark of marker 1 to -50 dBm.
Options:
R&S SMF-B20
Manual operation:
See
[:SOURce<hw>]:LFOutput:SWEep:VOLTage:MARKer<ch>:VSTate
<VState>
Activates the selected amplitude marker. If the marker is enabled, the signal level at
the MARKER output changes when the level line is reached.
Parameters:
<VState>
0 | 1 | OFF | ON
*RST:
0
SOURce Subsystem