Remote Control Commands
R&S
®
SMF100A
410
Operating Manual 1167.2319.02 ─ 12
Options:
Option R&S SMF-K28
:SENSe[:POWer]:SWEep:HCOPy:DEVice
<Device>
Defines the output device. The setting is fixed to FILE, i.e. the hardcopy is stored in a
file.
Parameters:
<Device>
FILE | PRINter
*RST:
FILE
Example:
SENS:SWE:HCOP:DEV FIL
selects output device file
Options:
Option R&S SMF-K28
Manual operation:
See
" Destination - Power Analysis "
:SENSe[:POWer]:SWEep:HCOPy:DEVice:LANGuage
<Language>
Selects the bitmap graphic format for the screenshot of the power analysis trace.
In addition, ASCII file format
*.csv
is offered. If file format
*.csv
is selected, the
trace data is saved as an ASCII file with comma separated values. It is also possible to
directly retrieve the data using command
SENS:SWE:HCOP:DATA?
.
Parameters:
<Language>
BMP | JPG | XPM | PNG | CSV
*RST:
BMP
Example:
SENS:SWE:HCOP:DEV:LANG BMP
selects output format
*.bmp
.
Options:
Option R&S SMF-K28
Manual operation:
See
:SENSe[:POWer]:SWEep:HCOPy:DEVice:LANGuage:CSV:DPOint
<DPoint>
Defines which character should be used as the decimal point of the values, either dot
or comma.
Parameters:
<DPoint>
DOT | COMMa
*RST:
DOT
Example:
SENS:SWE:HCOP:DEV:LANG CSV
selects output format
*.csv
.
SENS:SWE:HCOP:DEV:LANG:CSV:DPO DOT
selects character dot for being used as decimal point.
Options:
Option R&S SMF-K28
Manual operation:
See
" CSV Options Hardcopy- Power Analysis "
Power Sensor Measurement Subsystems