Remote Control Commands
R&S
®
SMF100A
405
Operating Manual 1167.2319.02 ─ 12
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:SRANge:STARt
<Start>
Sets the start frequency for the frequency power analysis with separate frequencies.
Parameters:
<Start>
integer
Range:
0 to 1E12
*RST:
1E6
Example:
SENS2:SWE:FREQ:SENS:SRAN:STAT ON
activates use of a separate frequency range for frequency ver-
sus power measurement for sensor 2.
SENS2:SWE:FREQ:SENS:STAR 2.0GHZ
sets a sweep start at 2 GHz irrespective of the current signal
generator frequency settings.
Options:
R&S SMF-K28
Manual operation:
See
" Min Frequency - Power Analysis "
:SENSe<ch>[:POWer]:SWEep:FREQuency[:SENSor]:SRANge:STOP
<Stop>
Sets the stop frequency for the frequency power analysis with separate frequencies.
Parameters:
<Stop>
integer
Range:
0 to 1E12
*RST:
10E6
Example:
SENS:SWE:FREQ:SENS2:SRAN:STAT ON
activates use of a separate frequency range for frequency ver-
sus power measurement.
SENS:SWE:FREQ:SENS2:STAR 2.0GHZ
sets a sweep start at 2 GHz irrespective of the current signal
generator frequency settings.
SENS:SWE:FREQ:SENS2:STOP 2.9GHZ
sets a sweep stop at 2.9 GHz irrespective of the current signal
generator frequency settings.
Options:
Option R&S SMF-K28
Manual operation:
See
" Max Frequency - Power Analysis "
:SENSe[:POWer]:SWEep:FREQuency:SPACing[:MODE]
<Mode>
Selects the spacing for the frequency power analysis.
Parameters:
<Mode>
LINear | LOGarithmic
*RST:
LINear
Example:
SENS:SWE:FREQ:SPAC:MODE LIN
sets linear spacing of the sweep
Power Sensor Measurement Subsystems