Remote Control Commands
R&S
®
SMF100A
398
Operating Manual 1167.2319.02 ─ 12
Manual operation:
See
:SENSe<ch>[:POWer]:FILTer:NSRatio
<NSRatio>
The command defines the noise content for fixed noise filter mode
(
:SENSe<[1]...3>:POWer:FILTer:TYPE NSRatio
). This value determines the
proportion of intrinsic noise in the measured result.
Parameters:
<NSRatio>
float
Range:
0.001 to 1
Increment: 0.001
*RST:
0.01
Example:
SENS1:FILT:TYPE NSR
selects fixed noise filter mode for the power sensor connected to
the SENSOR connector.
SENS1:FILT:NSR 0.2
sets a noise content of 0.2.
Manual operation:
See
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe
<MTime>
The command defines the timeout for fixed noise filter mode
(
:SENSe<[1]...3>:POWer:FILTer:TYPE NSRatio
). This value ensures limited
settling times.
Parameters:
<MTime>
float
Range:
1 to 999.99
Increment: 0.01
*RST:
4
Default unit: s
Example:
SENS1:FILT:TYPE NSR
selects fixed noise filter mode for the power sensor connected to
the SENSOR connector.
SENS1:FILT:NSR .2
sets a noise content of 0.2.
SENS1:FILT:NSR:MTIM 5
limits the settling time to 5 seconds
Manual operation:
See
:SENSe<ch>[:POWer]:FILTer:SONCe
The command activates the search for the optimum filter length for the current mea-
surement conditions. The found filter length can be retrieved with com-
mand
:SENSe:POWer:FILTer:LENGth:USER?
. This command is only available for
user filter mode (
:SENSe:POWer:FILTer:TYPE USER
).
Power Sensor Measurement Subsystems