![Rohde & Schwarz SMF-B1 Operating Manual Download Page 390](http://html1.mh-extra.com/html/rohde-and-schwarz/smf-b1/smf-b1_operating-manual_1477740390.webp)
Remote Control Commands
R&S
®
SMF100A
375
Operating Manual 1167.2319.02 ─ 12
Storing and Loading Current Settings
1. Store the current setting in an intermediate memory with the number 4. This setting
can be called using command
*RCL
and the associated number of the memory, for
example
*RCL 4
.
*SAV 4
2. To store the settings in a file in a specific directory, specify the complete path.
MMEM:STOR:STAT 4,"/var/user/test.savrcltxt"
3. To store the settings in a file in the default drive, set the default drive and specify
only the file name.
MMEM:CDIR '/var/user/'*SAV 4
MMEM:STOR:STAT 4,"test.savrcltxt"
4. Load the file
test.savrcltxt
in the
user
directory.
MMEM:LOAD:STAT 4,'/var/user/test.savrcltxt'
5. Activate the instrument setting of the file
test.savrcltxt
.
*RCL 4
Working with Files and Directories
1. Read out all files in the specified directory.
MMEM:CAT? '/usb/user'
Response:
127145265,175325184,"test,DIR,0","temp,DIR,0",
"readme.txt,ASC,1324","state.savrcltxt,STAT,5327",
"waveform.wv,BIN,2342"
the directory
/usb/user
contains the subdirectories
test
and
temp
as well as
the files
readme.txt
,
state.savrcltxt
and
waveform.wv
which have differ-
ent file types.
Tip:
To query only the subdirectories of the current or specified directory, perform:
MMEM:DCAT? '/usb/user'
Response:
'test', 'temp'
To query only the number of subdirectories in the current or specified directory,
perform:
MMEM:DCAT:LENG? '/usb/user'
Response:
2
2. To query the number of files in the current or specified directory, perform:
MMEM:CAT:LENG? '/usb/user'
Response:
3
3. Create a new subdirectory for mass memory storage in the specified directory.
MMEM:MDIR '/usb/new'
4. Copy the file
state
to a new file.
MMEM:COPY '/var/user/state.savrcltxt','/usb/new'
5. Rename the file
state
.
MMEMory Subsystem