Remote Control Commands
R&S
®
SMF100A
364
Operating Manual 1167.2319.02 ─ 12
Parameters:
<Border>
NORMal | SWAPped
NORMal
The instrument expects (with setting commands) and sends
(with queries) the least significant byte of each IEEE754 floating-
point number first and the most significant byte last.
SWAPped
The instrument expects (with setting commands) and sends
(with queries) the most significant byte of each IEEE754 floating-
point number first and the least significant byte last.
*RST:
NORMal
Example:
FORM:BORD SWAP
The data is transferred with the most significant bit first.
:FORMat[:DATA]
<Data>
Determines the data format which the R&S
SMF uses to return data. When data is
transferred from the control computer to the instrument, the instrument detects the data
format automatically. In this case, the value set here is irrelevant.
Parameters:
<Data>
ASCii | PACKed
ASCii
Numerical data is transferred as plain text separated by com-
mas.
PACKed
Numerical data is transferred as binary block data. The format
within the binary data depends on the command. The various
binary data formats are explained in the description of the
parameter types.
*RST:
ASCii
Example:
FORM ASC
The data is transferred as ASCII data.
:FORMat:SREGister
<Format>
Determines the numerical format which is returned when the status registers are quer-
ied.
FORMat Subsystem