Instrument Function
R&S
®
SMF100A
198
Operating Manual 1167.2319.02 ─ 12
Use the "Setup >"
Chapter 5.2.3.5, "NRP-Z Info/Update"
the sensor software.
Remote command:
SENSe<ch>[:POWer]:STATus[:DEVice]?
Type
Indicates the type and the serial number of the connected R&S NRP-Z power sensor.
The sensor type is automatically detected.
Remote command:
State
Activates/deactivates level measurement by the power sensor.
The local state is set with the
INIT
command. Switching the local state off enhances
the measurement performance.
In remote control, the sensors are set up using the
SENSe
commands. The remote
measurement is triggered by the
READ
query which also provides the measurement
results. The state is not influenced by these commands, measurements results can be
retrieved with local State on or off.
The sensor is selected by suffix 1, 2, 3 or 4 in key word
SENSe
or
READ
of the com-
mand header.
Suffix 1 denotes the sensor connected to SENSOR, suffix 2 the one at the first USB
interface, and suffix 3 and 4 are assigned to the sensors at the following USB interfa-
ces. The suffix is identical to the index which is assigned automatically to each sensor
upon connection.
In order to detect all connected sensors the state of all four connectors (i.e. SENsor1/
SENSor2/SENSor3/SENSor4) must be checked.
To query the availability of a sensor at a given connector, use the command
SENSe<ch>[:POWer]:STATus[:DEVice]?
Remote command:
:INITiate<ch>[:POWer]:CONTinuous
Level (Peak)
With certain power sensors only, for example R&S NRP-Z81.
Indicates the measured peak level value with the selected unit.
Remote command:
Level (Avg.)
Indicates the measured level value with the selected unit.
Remote command:
Unit
Selects the unit used for result display.
RF Level