Operating Manual 1411.4060.32 ─ 04
STATus:OPERation:NTRansition
<Ntransition>
The command sets the bits of the NTRansition part of the STATus:OPERation register.
If a bit is set, a transition from 1 to 0 in the condition part causes an entry to be made in
the EVENt part of the register. The disappearance of an event in the hardware is thus
registered, e.g. the end of an adjustment.
Parameters:
<Ntransition>
string
Example:
STAT:OPER:NTR 0
a transition from 1 to 0 in the condition part of the Status:Operation
register does not cause an entry to be made in the EVENt part.
Usage:
SCPI confirmed
STATus:OPERation:PTRansition
<Ptransition>
The command sets the bits of the PTRansition part of the STATus:OPERation register.
If a bit is set, a transition from 0 to 1 in the condition part causes an entry to be made in
the EVENt part of the register. A new event in the hardware is thus registered, e.g. the
start of an adjustment.
Parameters:
<Ptransition>
string
Example:
STAT:OPER:PTR 32767
all transitions from 0 to 1 in the condition part of the Status:Oper-
ation register cause an entry to be made in the EVENt part.
Usage:
SCPI confirmed
STATus:PRESet
<Preset>
The command resets the status registers. All PTRansition parts are set to FFFFh (32767),
i.e. all transitions from 0 to 1 are detected. All NTRansition parts are set to 0, i.e. a tran-
sition from 1 to 0 in a CONDition bit is not detected. The ENABle parts of STATus:OPER-
ation and STATus:QUEStionable are set to 0, i.e. all events in these registers are not
passed on.
Parameters:
<Preset>
string
Example:
STAT:PRES
resets the status registers.
Usage:
Event
SCPI confirmed
STATus Subsystem