Remote Control Commands
R&S
®
SMC100A
296
Operating Manual 1411.4060.32 ─ 04
[:SOURce<hw>]:SWEep:POWer:EXECute
The command triggers a sweep.
The command is only valid for sweep mode Single (
SOURce:SWEep:POWer:MODE
SINGle
). The command corresponds to the manual-control command "Execute Single
Sweep".
Example:
SOURce:SWEep:POWer:MODE SINGle
sets the single cycle mode of the level sweep.
SWE:POW:EXEC
triggers a level sweep at the RF output.
Usage:
Event
[:SOURce<hw>]:SWEep:POWer:MODE
<Mode>
The command sets the cycle mode of the level sweep.
Parameters:
<Mode>
AUTO | MANual | STEP
AUTO
Each trigger triggers exactly one complete sweep.
MANual
The trigger system is not active. Each level step of the sweep is
triggered individually, either by varying the "Current Level" value
using the rotary knob under manual control or by means of a
POW:MAN
command under remote control.
With manual control, the level increases or decreases (depending
on the direction of the rotary encoder) by the value specified under
SOUR:SWE:POW:STEP
. With remote control, the level increases
by the value specified under
SWEep:POW:STEP
which each
sent
:POW:MAN
command, irrespective the value entered there.
STEP
Each trigger triggers one sweep step only. The level increases by
the value entered under
:SWEep:POWer:STEP
.
*RST:
AUTO
Example:
SWE:POW:MODE AUTO
selects Mode Auto for a level sweep at RF output.
[:SOURce<hw>]:SWEep:POWer:POINts
<Points>
The command sets the number of steps in a level sweep. The command is linked to the
command
:SWEep:POWer:STEP
as follows:
POINts = ((STOP - STARt) / STEP:LOG) + 1
If
POINTs
changes, the value of
STEP
is adjusted. The
STARt
and
STOP
value is retained.
SOURce Subsystem