Remote Control Commands
R&S
®
SMC100A
285
Operating Manual 1411.4060.32 ─ 04
Example:
PULM:MODE DOUB
enables double pulse generation.
[:SOURce<hw>]:PULM:PERiod
<Period>
The command sets the period of the generated pulse. The period determines the repe-
tition frequency of the internal signal.
Parameters:
<Period>
float
Range:
5us | 20ns to 100 s
Increment: 1us | 20ns
*RST:
1ms
Example:
PULM:PER 220 us
the pulse period is 220 us.
[:SOURce<hw>]:PULM:POLarity
<Polarity>
The command sets the polarity between modulating and modulated signal. This com-
mand is only effective for an external modulation signal.
Parameters:
<Polarity>
NORMal | INVerted
NORMal
The RF signal is suppressed during the pulse pause.
INVerted
The RF signal is suppressed during the pulse.
*RST:
NORMal
Example:
PULM:SOUR EXT
selects the external modulation source.
Example:
PULM:POL INV
selects inverted polarity.
[:SOURce<hw>]:PULM:SOURce
<Source>
The command selects the source for pulse modulation.
SOURce Subsystem