Remote Control Commands
R&S
®
SMC100A
283
Operating Manual 1411.4060.32 ─ 04
Example:
POW:STOP 3
sets the stop level for the level sweep to 3 dBm for RF output A.
Usage:
SCPI confirmed
7.12.12 SOURce:PULM Subsystem
This subsystem contains the commands for setting the pulse modulation.
The LF generator is used as the internal modulation source. The pulse frequency of the
internal rectangular signal is therefore set in the
SOURce:LFOutput
subsystem.
The external signal is input at the PULSE EXT connector. The connector can be used as
trigger input for internal pulse modulation. The polarity and input impedance of the con-
nector can be selected. The pulse modulation signal is output at the PULSE VIDEO con-
nector.
........................................................................................283
............................................................................283
...........................................................................284
...........................................................................284
........................................................................................284
.......................................................................................285
.....................................................................................285
.....................................................................................285
........................................................................................286
................................................286
............................................................286
...........................................................................287
........................................................................................287
[:SOURce<hw>]:PULM:DELay
<Delay>
The command sets the pulse delay. The pulse delay determines the time that elapses
after a trigger event before pulse modulation starts. The pulse delay is not effective for
double pulse generation.
Parameters:
<Delay>
float
Range:
10 ns to 100 s
Increment: 20 ns
*RST:
20 ns
Default unit: ns
Example:
PULM:DEL 13 us
13 us elapse after a trigger before the first pulse is generated.
[:SOURce<hw>]:PULM:DOUBle:DELay
<Delay>
The command sets the delay from the start of the first pulse to the start of the second
pulse.
SOURce Subsystem