Remote Control Commands
R&S
®
SMC100A
278
Operating Manual 1411.4060.32 ─ 04
Example:
SOURce:POWer:LEVel:IMMediate:OFFSet -10
or
POW:OFFS 10
sets the RF level offset to 10 dB
Usage:
SCPI confirmed
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:RCL
<Rcl>
The command determines whether the current level is retained or whether the stored
level setting is adopted when an instrument configuration is loaded.
*RST
does not affect this setting.
Parameters:
<Rcl>
INCLude | EXCLude
INCLude
The stored level is also loaded when a stored instrument config-
uration is loaded.
EXCLude
The RF level is not loaded when a stored instrument configuration
is loaded. The current level is retained.
*RST:
no default
Example:
POW:RCL INCL
the stored level is set if the Recall command is called.
Usage:
SCPI confirmed
[:SOURce<hw>]:POWer:LIMit[:AMPLitude]
<Amplitude>
The command limits the maximum RF output level in CW and SWEEP mode. It does not
influence the "Level" display or the response to the
POW?
query command.
The value is not affected by an instrument preset (PRESET key),
*RST
and the Save/
Recall function. This parameter is influenced only by the factory preset (
SYST:FPR
) and
its factory value is equal to the upper limit.
Parameters:
<Amplitude>
float
Minimum level ... Maximum level
The value range for the level setting varies according to the instru-
ment model.
The values are given in the data sheet.
Increment: 0.01 dB
*RST:
30 dBm
Default unit: dBm
SOURce Subsystem