Remote Control Commands
R&S
®
SMC100A
276
Operating Manual 1411.4060.32 ─ 04
Parameters:
<State>
ON | OFF | AUTO
ON
Internal level control is permanently activated.
OFF
Internal level control is deactivated; Sample & Hold mode is acti-
vated.
AUTO
Internal level control is activated/deactivated automatically
depending on the operating state.
*RST:
AUTO
Example:
POW:ALC ON
activates automatic level control for RF output A.
[:SOURce<hw>]:POWer:ATTenuation:RFOFf:MODE
<Mode>
The command selects the attenuator mode, when the RF signal is switched off.
The setting of the RF OFF mode is not affected by an instrument preset (PRESET key),
*RST
and the "Save/Recall" function. This parameter is influenced only by the
.
Parameters:
<Mode>
UNCHanged | FATTenuated
UNCHanged
Freezes the setting of the attenuator when RF is switched off. The
attenuator is only activated when RF is switched on.
This setting recommended if a constant VSWR (
V
oltage
S
tanding
W
ave
R
atio) is required.
FATTenuated
Sets attenuation to maximum when the RF signal is switched off.
This setting is recommended for applications that require a high
level of noise suppression.
*RST:
FATTenuated
Example:
SOUR:POW:ATT:RFOF:MODE FATT
sets the RF OFF attenuator to maximum.
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude]
<Amplitude>
The command sets the RF output level in CW mode. In addition to numerical values, it
is also possible to specify
UP
and
DOWN
. The level is then increased or decreased by the
value specified under
[SOURce<hw>:]POWer:STEP
.
As with the
Level
value entered in the header, the
OFFSet
value is also taken into con-
sideration with this command.
SOURce Subsystem