Remote Control Commands
R&S
®
SMC100A
270
Operating Manual 1411.4060.32 ─ 04
[:SOURce]:LFOutput:VOLTage
<Voltage>
The command sets the output voltage of the LF output.
Parameters:
<Voltage>
float
Range:
0 to 4
Increment: 0.001
*RST:
1
Default unit: V
Example:
LFO:VOLT 3 V
sets the voltage of the LF output to 3 V.
7.12.7 SOURce:MODulation Subsystem
This subsystem contains the command for switching on/off all modulations.
[:SOURce<hw>]:MODulation[:ALL]:STATe
<State>
The command switches the modulations on and off.
The command
SOUR:MOD:ALL:STAT OFF
switches all modulations off. A subsequent
command
SOUR:MOD:ALL:STAT ON
restores the status that was active before the last
switch-off. "MOD OFF" is displayed in the info line of the header next to the "Level" field.
Parameters:
<State>
0 | 1 | OFF | ON
Example:
MOD:STAT OFF
switches off all modulations.
7.12.8 SOURce:PGEN Subsystem
This subsystem contains the commands for setting the pulse generator.
........................................................................................270
[:SOURce<hw>]:PGEN:STATe
<State>
The command activates/deactivates the output of the video/sync signal at the PULSE
VIDEO connector at the rear of the instrument.
The signal output and the pulse generator are automatically switched on with activation
of pulse modulation if pulse generator is selected as modulation source. The signal output
can be switched off subsequently.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
OFF
SOURce Subsystem