Remote Control Commands
R&S
®
SMC100A
209
Operating Manual 1411.4060.32 ─ 04
*STB?
STatus Byte query
Reads the contents of the status byte in decimal form.
Usage:
Query only
*TRG
TRiGger
Triggers all actions waiting for a trigger event. In particular, *TRG generates a manual
trigger signal (Manual Trigger). This common command complements the commands of
the
TRIGger
subsystem.
Usage:
Event
*TST?
self TeST query
Triggers selftests of the instrument and returns an error code in decimal form (see Service
Manual supplied with the instrument). "0" indicates no errors occured.
Usage:
Query only
*WAI
WAIt to continue
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
).
Usage:
Event
7.2 Preset Commands
The preset commands are not bundled in one subsystem. Therefore, they are listed sep-
arately in this section. In addition, a specific preset command is provided for each digital
standard and for the fader. These specific commands are described in the associated
subsystems.
Four presetting actions are available:
●
Activating the default state of all internal instrument functions (
Functions that concern the integration of the instrument into a measurement setup
are not changed, e.g. GPIB address or reference oscillator source settings.
●
Activating the preset state of the parameters related to the selected signal path (
Preset Commands