Instrument Function
R&S
®
SMC100A
144
Operating Manual 1411.4060.32 ─ 04
"Extern Start/
Stop"
Sets an automatically repeated sweep cycle that is started, stopped and
restartet by subsequent external trigger events.
The first external trigger signal starts the sweep (Start).
The next external trigger signal stops the sweep at the current fre-
quency (Stop).
The third external trigger signal starts the sweep at the start frequency
(Start).
If a different sweep mode was activated prior to the "Extern Start/
Stop" mode, the current sweep is stopped and the "Extern Start/Stop"
sweep starts at the start frequency when triggered.
The "Reset" button resets the sweep to the start level.
Refer to the description of the rear panel for information about the con-
nectors for external trigger signal input (see section "Legend for Rear
Panel View").
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR EAUT
SOUR:POW:MODE SWE
(External trigger)
SCPI command:
Reset Sweep - Level Sweep
Resets the sweep. The start level is set and the next sweep starts from there.
SCPI command:
Execute Single Sweep - Level Sweep
Triggers the sweep manually. A manual sweep can only be triggered if "Mode Single" is
selected.
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR SING
SOUR:POW:MODE SWE
SOUR:SWE:EXEC
SCPI command:
Start Level - Level Sweep
Sets the start level.
SCPI command:
RF Signal - RF Block