Instrument Function
R&S
®
SMC100A
138
Operating Manual 1411.4060.32 ─ 04
"Extern Start/
Stop"
Sets an automaticly repeated sweep cycle that is started, stopped and
restartet by subsequent external trigger events.
The first external trigger signal starts the sweep (Start).
The next external trigger signal stops the sweep at the current fre-
quency (Stop).
The third external trigger signal starts the sweep at the start frequency
(Start).
If a different sweep mode was activated prior to the "Extern Start/
Stop" mode, the current sweep is stopped and the "Extern Start/Stop"
sweep starts at the start frequency when triggered.
The "Reset" button resets the sweep to the start frequency.
Refer to the description of the rear panel for information about the con-
nectors for external trigger signal input.
The external trigger signal is input at the rear of the instrument (BNC
connector INST TRIG).
Example:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR EAUT
SOUR:FREQ:MODE SWE
(External trigger)
SCPI command:
Execute Single Sweep - Frequency Sweep
Triggers the sweep manually. A manual sweep can only be triggered if "Mode Single" is
selected.
SCPI command:
Reset Sweep - Frequency Sweep
Resets the sweep. The start frequency is set and the next sweep starts from there.
SCPI command:
Start Freq - Frequency Sweep
Sets the start frequency.
SCPI command:
Stop Freq - Frequency Sweep
Sets the stop frequency.
SCPI command: