![Rohde & Schwarz SMC100A Operating Manual Download Page 152](http://html.mh-extra.com/html/rohde-and-schwarz/smc100a/smc100a_operating-manual_1477935152.webp)
Instrument Function
R&S
®
SMC100A
135
Operating Manual 1411.4060.32 ─ 04
RF Frequency Sweep Settings
State - Frequency Sweep
Activates/deactivates RF sweep mode.
Note:
Activating the RF-Sweep mode automatically disables other sweep modes.
SCPI command:
Mode - Frequency Sweep
Selects the sweep instrument operating mode and the sweep mode.
"Auto"
Sets an automatic repeated sweep cycle. If a different sweep mode was
activated prior to the "Auto" mode, the cycle continues from the current
sweep setting.
The "Reset Sweep" button resets the sweep to the start frequency.
Example:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR AUTO
SOUR:FREQ:MODE SWE
RF Signal - RF Block