
Remote Control Commands
R&S
®
SMBV100B
896
User Manual 1178.4460.02 ─ 03
Example:
SYST:DLOC ON
Activates the display lock. The instrument cannot be operated
via the display until it has been enabled with
SYST:DLOC OFF
.
Manual operation:
See
:SYSTem:KLOCk
<State>
Disables the front panel keyboard of the instrument including the Local key.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
n.a. (factory preset: 0)
Example:
SYST:KLOC ON
Locks the front panel and external controls.
To enable the controls, send
SYST:KLOC OFF
.
Manual operation:
See
:SYSTem:SAV
<Pathname>
Saves the current instrument settings to a file with defined filename.
Setting parameters:
<Pathname>
string
Example:
SYSTem:SAV
"/var/user/Instrument_settings.svrcltxt"
Usage:
Setting only
:SYSTem:RCL
<Pathname>
Loads a file with previously saved instrument settings.
Setting parameters:
<Pathname>
string
Example:
SYSTem:RCL
"/var/user/Instrument_settings.svrcltxt"
Usage:
Setting only
:SYSTem:PROTect<ch>[:STATe]
<State>[, <Key>]
Activates and deactivates the specified protection level.
Suffix:
<ch>
.
Indicates the protection level.
See also
SYSTem Subsystem