Remote Control Commands
R&S
®
SMBV100B
716
User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:BB:DM:SOURce
<Source>
Selects the data source.
Parameters:
<Source>
ZERO | ONE | PRBS | PATTern | DLISt
A sequence of 0 or 1, a pseudo-random sequence with different
length, a pattern, or a data list.
*RST:
PRBS
Example:
See
Example "Performing general tasks"
Manual operation:
See
[:SOURce<hw>]:BB:DM:PATTern
<Pattern>, <BitCount>
Selects the data pattern for the internal data source.
Parameters:
<Pattern>
numeric
*RST:
#H0
<BitCount>
integer
Range:
1 to 64
*RST:
1
Example:
SOURce:BB:DM:SOURce PATT
SOURce:BB:DM:PATTern #B01110111010101010,17
Generates the user-defined sequence of 0/1 data.
Manual operation:
See
[:SOURce<hw>]:BB:DM:PRBS[:LENGth]
<Length>
Defines the length of the pseudo-random sequence in accordance with the following
equation:
Length = (2^
Length
) - 1
Parameters:
<Length>
9 | 11 | 15 | 16 | 20 | 21 | 23 | PN9 | PN11 | PN15 | PN16 | PN20 |
PN21 | PN23
*RST:
9
Example:
See
Example "Performing general tasks"
Manual operation:
See
SOURce Subsystem