Remote Control Commands
R&S
®
SMBV100B
710
User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:BBIN:FOFFset
<FOffset>
[:SOURce<hw>]:BB:FOFFset
<FOffset>
Sets a frequency offset for the internal/external baseband signal. The offset affects the
generated baseband signal.
Parameters:
<FOffset>
float
Range:
depends on the installed options
Increment: 0.01
*RST:
0
Default unit: Hz
E.g. -60 MHz to +60 MHz (base unit)
Example:
SOURce1:BB:FOFFset 2MHZ
Manual operation:
See
[:SOURce<hw>]:BBIN:POFFset
<POffset>
Sets the relative phase offset for the external baseband signal.
Parameters:
<POffset>
float
Range:
-999.99 to 999.99
Increment: 0.01
*RST:
0
Default unit: DEG
Example:
SOURce1:BBIN:POFFset 0.5
Manual operation:
See
[:SOURce<hw>]:BB:POFFset
<POffset>
Sets the relative phase offset for the selected baseband signal.
Parameters:
<POffset>
float
Range:
0 to 359.9
Increment: 0.01
*RST:
0
Default unit: DEG
Example:
SOURce1:BB:POFFset 0.5DEG
Manual operation:
See
[:SOURce<hw>]:BB:ROUTe
<Route>
[:SOURce<hw>]:BBIN:ROUTe
<Route>
Selects the signal route for the internal/external baseband signal.