Remote Control Commands
R&S
®
SMBV100B
694
User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:AM:RATio
<Ratio>
Sets the deviation ratio (path#2 to path#1) in percent.
Parameters:
<Ratio>
float
Range:
0 to 100
Increment: 0.01
*RST:
100
Example:
See
Example "Creating an amplitude modulated RF signal"
Manual operation:
See
[:SOURce<hw>]:AM<ch>:SENSitivity
<Sensitivity>
Sets the sensitivity of the external signal source for amplitude modulation in %/V.
Parameters:
<Sensitivity>
float
Range:
0 to 100
Increment: 0.1
*RST:
30
Example:
See
Example "Using an external signal source"
Manual operation:
See
12.16.3.3
SOURce:FM Subsystem
Option: R&S
SMBVB-K720
The FM subsystem contains the commands for setting the frequency modulation.
Example: Creating a frequency modulated RF signal
Using the internal LF generator, the following command sequence configures a fre-
quency modulated signal.
// Reset the instrument to start from a defined state
*RST
// Set RF frequency and amplitude
SOURce1:FREQuency:CW 6000000000
SOURce1:POWer:LEVel:IMMediate:AMPLitude -25
// Configure the modulation signal
SOURce1:LFOutput1:SHAPe SINE
SOURce1:LFOutput1:FREQuency 20000
// Configure the frequency modulation settings and switch FM on
SOURce1:FM1:SOURce LF1
SOURce Subsystem