Network Operation and Remote Control
R&S
®
SMBV100B
572
User Manual 1178.4460.02 ─ 03
Parameter
Value
ICMP ping
Enabled
Password for LAN configuration
LxiWebIfc
The LAN reset also resets the following parameters for the vector signal generator:
Parameter
Hostname
Instrument-specific hostname
Description
Vector signal generator
Negotiation
Auto detect
VXI-11 discovery
Enabled
11.5.2
LXI Browser Settings
To access the instrument via the web browser:
► In the address field of the browser on your PC, type the instrument's host name or
IP address, for example
http://10.113.1.151
.
Note:
Do not add the missing zeros in the IP address, while opening the instru-
ment homepage.
The instrument homepage (welcome page) opens.
The navigation pane of the browser interface contains the following elements:
● "LXI"
–
"Home" opens the instrument homepage.
The homepage displays the device information required by the LXI stan-
dard, including the VISA resource string in read-only format.
–
"Device Indicator" activates or deactivates the LXI status indication.
When activated, the LXI LEDs flash, both in the browser dialog and in the
LXI dialog of the connected instrument, see
. A green LXI status
symbol indicates that a LAN connection has been established; a red sym-
bol indicates that no LAN cable is connected.
–
"Lan Configuration" allows you to configure LAN parameters and to initiate
a ping, see
–
"Status" displays information about the LXI status of the instrument.
–
"Utilities" provides access to the LXI event log functionality required by the
LXI standard.
● "Instrument Control"
–
"Web Control" provides remote access to the instrument, see
remote control via the LXI web browser"
● "Diagnostics"
–
"SCPI Remote Trace" records messages exchanged via the remote control
interface, see
● "Help"
–
"Glossary" explains terms related to the LXI standard.
LXI Settings