Multi-Instrument Setups
R&S
®
SMBV100B
448
User Manual 1178.4460.02 ─ 03
External instruments connected to the R&S
) whenever:
●
The connection to the external instrument is established or terminated ("Connect/
Disconnect All Remote" or "Rem. Conn > On/Off")
●
The R&S
SMBV100B is preset ([Preset]).
Note:
An active "Rem. Conn." and an "RF State > On" do not mean that an I/Q stream
is mapped to this connector.
Use the "System Configuration > External Rf and I/Q > Display > Mapped Connectors"
filter to list only the connectors with a mapped I/Q stream.
Remote command:
:SCONfiguration:EXTernal:DIGital:RCONnection:STATe?
:SCONfiguration:EXTernal:DIGital:REMote:DISConnect
:SCONfiguration:EXTernal:IQOutput:RCONnection:STATe?
:SCONfiguration:EXTernal:IQOutput:REMote:DISConnect
:SCONfiguration:EXTernal:DIGital:REMote:INFO?
:SCONfiguration:EXTernal:IQOutput:REMote:INFO?
Instrument Name
Displays useful information on the connected external instrument, like the instrument
designation and the RF path the RF settings are coupled to.
Remote command:
:SCONfiguration:EXTernal:DIGital:INAMe?
:SCONfiguration:EXTernal:IQOutput:INAMe?
:SCONfiguration:EXTernal:DIGital:REMote:INFO?
:SCONfiguration:EXTernal:IQOutput:REMote:INFO?
RF Couple
If enabled, this parameter couples all major RF setting (like the "RF Frequency", the
"RF Level" and the "RF State") of the external instrument to the R&S
SMBV100B.
The R&S
SMBV100B acts as a "master", i.e the external instruments synchronize their
settings and the RF output state to the R&S
SMBV100B. You can apply delta fre-
quency and level offset to the signal.
Remote command:
:SCONfiguration:EXTernal:DIGital:RF:COUPling
:SCONfiguration:EXTernal:IQOutput:RF:COUPling
(Delta) RF Frequency/ RF Level
In uncoupled mode, sets the RF frequency and RF level of the external instrument.
With enabled "RF Coupled" parameter, applies a frequency and level offset. The RF
frequency of the external instrument is calculated as follows:
RF
ExtInstr
= RF
Instrument
+ Delta
Freq
Where both the RF frequency and the frequency offset are applied at the external
instrument.
Remote command:
:SCONfiguration:EXTernal:DIGital:RF:FREQuency
:SCONfiguration:EXTernal:IQOutput:RF:FREQuency
System Configuration Settings