Configuring the RF Signal
R&S
®
SMBV100B
323
User Manual 1178.4460.02 ─ 03
Figure 6-13: Sweep signal sawtooth shape
1 = Start value
2 = Stop value
3 = Step size
●
Triangle
The sweep sequence resembles a triangle. One sweep runs from start to stop
value frequency and back. Each subsequent sweep starts at the start value.
Figure 6-14: Sweep signal triangle shape
1 = Start value
2 = Stop value
3 = Step size
The current frequency or level of an RF frequency, level or LF frequency sweep is
determined by the shape (e.g. sawtooth or triangle) and the spacing (linear or logarith-
mic), and the step size.
6.10.3
About List Mode
The list mode is used to generate the RF signal based on a set of predefined fre-
quency and amplitude value pairs, with individual step times. You can define the values
arbitrarily, in any order and varying step sizes, within the entire configurable value
range of the instrument.
Varying the RF Signal in List or Sweep Mode