Adding Noise and Predistortions, and Impairing the Signal
R&S
®
SMBV100B
259
User Manual 1178.4460.02 ─ 03
Figure 4-3: AWGN: Understanding the displayed information
1 =
derived from the selected "Symbol Rate = 3.840
Msym/s"
2 = Selected
3 =
= Status bar > Level =
30
dBm
4 =
calculated form the "Carrier Power" and the "Carrier/Noise Ratio" = -30
dBm - 10
dB =
-40
dBm
5 =
= power at the RF output; i.e. the power a connected power meter would measure
6 =
Minimum Noise/System Bandwidth Ratio
MHz*3 =
11.520
MHz
Use the built-in graphical signal monitoring function of the instrument to display the sig-
nal characteristics in real-time, see
"To observe the effect of enabled additive noise
Application fields
Typically, the noise generator is required for the following tests:
●
In receiver sensibility tests with predefined SNR of the receiver, see
the AWGN for receiver sensibility tests with predefined SNR of the receiver"
●
In bit-error or block-error measurements, depending on the set signal-to-noise ratio
(SNR)
●
Whenever a pure noise signal is required
4.1.3
AWGN Settings
The "AWGN" block provides access to the settings for the AWGN generator.
How to access and configure the "AWGN Settings"
1. Select "Block Diagram > AWGN".
Adding Noise to the Signal