Configuring the Internal Baseband Source
R&S
®
SMBV100B
214
User Manual 1178.4460.02 ─ 03
"Next Segment Seamless"
If all segments have the same sample rate, the new segment is not
output until the whole of the current segment has been output (wrap
around), i.e. the signal transition is seamless.
"Sequencer"
The waveform files are processed according to the order and repeti-
tion cycles defined in the special sequencing list file (
*.wvs
)
assigned to the multi-segment waveform file.
This “play list” is defined in the
Multi-Segment Waveform Sequencing
dialog.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:TRIGger:SMODe
Next Segment Source
Determines whether the next segment is switched upon receiving of an external trigger
signal or upon executing of software next segment trigger (see
).
This parameter is disabled, if a sequencing play list is used.
"Internal"
Switch over to any segment is triggered manually with the "Execute
Next Segment" function.
"Global Next Segment"
The trigger event for switching to the next segment is input as config-
ured on one of the User x connectors.
See
Chapter 10.2, "Configuring the Global Connectors"
Remote command:
[:SOURce<hw>]:BB:ARBitrary:WSEGment:NEXT:SOURce
Level Mode
Sets how the instrument levels the individual segments in a multi-segment signal. The
parameter is important, if waveforms with different crest factors are used.
"Highest"
All segments are output at the highest available crest factor.
"Unchanged"
Each segment is output with the crest factor defined in its waveform
file.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:WSEGment:LMODe
Execute Next Segment
For "Next Segment Source = Internal", triggers manually switch over to the subsequent
segment in the multi-segment file. The subsequent segment can be any segment
within the multi-segment file and is defined by the parameter "Segment"
This parameter is disabled, if a sequencing play list is enabled.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:WSEGment:NEXT:EXECute
Sequencing List
Opens a dialog box for defining the “play lists”, see
Chapter 3.7.3.4, "Multi-Segment
Generating Multi-Segment Waveform Files