Remote Control Commands
R&S
®
SMBV100A
790
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Amplitude>
float
Minimum level ... Maximum level
The value range for the level setting varies according to the
instrument model.
The values are given in the data sheet.
Increment: 0.01
*RST:
n.a. (factory preset: 30)
Example:
SOURce:POWer:LIMit:AMPLitude 10
or
:POW:LIM 10
limits the RF level to m10 dBm.
Manual operation:
See
[:SOURce<hw>]:POWer:LMODe
<LevMode>
Sets the RF level mode.
Parameters:
<LevMode>
NORMal | LOWNoise | LOWDistortion
NORMal
The RF signal is output in the standard values of the instrument.
LOWNoise
A very low noise sinewave signal is output.
LOWDistortion
A very pure sinewave signal is output.
*RST:
NORMal
Example:
POW:LMODE LOWD
sets the LOWDistortion mode. The instrument reduces distor-
tions of the RF signal to a minimum.
Manual operation:
See
[:SOURce<hw>]:POWer:MANual
<Manual>
In Sweep mode
(:SOUR:POW:MODE SWE
) the command sets the level for the next
sweep step in the Step sweep mode (
:SOUR:SWE:POW:MODE MAN
). Here only level
values between the settings
[:SOUR]:POW:STAR
and
[:SOUR]:POW:STOP
are per-
mitted. Each sweep step is triggered by a separate
:SOUR:POW:MAN
command.
As with the "Level" value entered in the "RF Level" menu, the
OFFSet
value is also
taken into consideration with this command.
The specified value range is therefore only effective if
:SOURce:POWer:OFFSet
is set
to 0. The value range for other
OFFset
values can be calculated using the following
formula:
Minimum level +
OFFSet
... Maximum level +
OFFSet
SOURce Subsystem