Remote Control Commands
R&S
®
SMBV100A
750
Operating Manual 1176.8016.02 ─ 17
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
IQ:OUTP:BIAS:ANAL:COUP ON
activates I/Q bias coupling.
Manual operation:
See
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:BIAS:I
<I>
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:BIAS:Q
<Q>
Specifies the amplifier bias of the respective I/Q- component.
Parameters:
<Q>
float
Range:
-3.6 to 3.6
Increment: 0.002
*RST:
0
Example:
IQ:OUTP:ANAL:BIAS:Q 0.1 V
sets a bias of 100 mV for the Q-signal.
Manual operation:
See
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:OFFSet:I
<I>
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:OFFSet:Q
<Q>
Valid only for output type Differential (
IQ:OUTP:ANAL:TYPE DIFF
).
Sets an offset between the inverting and non-inverting output.
Parameters:
<Q>
float
Range:
-0.3 to 0.3
Increment: 1E-4
*RST:
0
Example:
IQ:OUTP:ANAL:TYPE DIFF
the I/Q-signal components are output differential
:IQ:OUTP:ANAL:OFFS:Q 0.05 V
sets an offset of 50 mV for the I-signal.
Manual operation:
See
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:BIAS:OPTimization[:RF]:STATe
<State>
Activates/deactivates optimization of level accuracy and imbalance for RF Output.
Level accuracy and imbalance can be either optimized for RF output (active) or for dif-
ferential outputs (not active).
SOURce Subsystem