Remote Control Commands
R&S
®
SMBV100A
691
Operating Manual 1176.8016.02 ─ 17
Parameters:
<Q>
float
Range:
-10 to 10
Increment: 0.01
*RST:
0
Default unit: PCT
Example:
BB:IMP:LEAK:Q 3 PCT
sets the leakage for the Q-component to 3 percent.
Manual operation:
See
"I/Q Offset - Digital Impairments"
[:SOURce<hw>]:BB:IMPairment:QUADrature[:ANGLe]
<Angle>
This command sets the quadrature offset for the digital I/Q signal.
Parameters:
<Angle>
float
Range:
-10 to 10
Increment: 0.01
*RST:
0
Default unit: DEG
Example:
BB:IMP:QUAD:ANGL -5DEG
sets the quadrature offset to -5 degrees.
Manual operation:
See
"Quadrature Offset - Digital Impairments"
[:SOURce<hw>]:BB:IMPairment:SKEW
<Skew>
Sets the time offset between the I and Q vectors.
In an I/Q modulator without I/Q skew, the I and Q vectors are aligned to the marker.
With an I/Q skew, both vectors are shifted relative to the marker so that the offset of
each of the vectors to the marker will be the half of the I/Q skew value.
A positive I/Q skew means that the Q vector delays relative to the I vector and vice
versa.
Parameters:
<Skew>
float
Range:
-500E-9 to 500E-9
Increment: 1E-12
*RST:
0
Example:
BB:IMP:SKEW 410.0E-9
sets the I/Q skew.
Manual operation:
See
"I/Q Skew - Digital Impairments"
SOURce Subsystem